Targets

- 820-Liquid-lanthanum. To be filled today in the Glove box (Bernard, Valantina)


- #858-UC-MK1-Ta Gold mass marker issue investigated. 3 possible causes: 
    - Mass marker capillary detached from back (to check photos and voltages on other mass markers (they are all somewhat connected through the return))
    - inserting the sample into the MM by wire. wire might stick to the sample and be pulled back when removing the wire. Sample might be located on the wring side of the crimp. ()
    - Sample came out completely (not likely)


- #534-Sn Outgassing / leak when heating. Decided finally to mount the spare #567-Sn . Spare was tested at offline 1 prior to installation at GPS.


- Target for winter physics #835-UC was irradiated at the Medicis position and now cannot be pumped at the MEDICIS frontend. Suggest to couple to HRS or GPS to see if it is Frontend related or indeed a degradation of seals in the Target unit. (This is now the third case)
- Suggest to perform in parallel a heat test with a base from this series. Put the same electrical power through a target container and measure temperatures on all feedthroughs and critical locations.

 

Bernard loading the #820 with lanthanum.