MEDICIS
Last Week
- VD-5 TaC testing in the framework of Patricija's PhD.
- Fluoride molecules with CF4 injection.
- Noticable CO+ outgassing from the target.
- Indirect irradiation of TaC for Tb run during the week-end.
- Small delay placing the target on low trolley - trolley was not fully at back position, however switch indicated good position.
This Week
- VD5 TaC testing in the framework of Patricija's PhD.
- The temperature sensor on target feedthrough did not show increased temperature on the target (>116 oC). Cooling - to be also tested during direct irradiation on thursday.
- Fluoride molecules with CF4 injection.
- Direct irradiation of TaC for Tb run → target will be sent on the high trolley on Thursday afternoon 14:30.
- Tb collection continued - If enough activity >1 MBq collected, sample will be shipped to NPL, UK.
Next week
- Tb-run (Patricija PhD) continued.
- ThC loading inside the target with target testing and pumping on MEDICIS FE.
Class A
- New firefighter training yesterday morning
- Visit from Cheryl Saban tomorrow
- Visit from the french upper chamber senate thurs
- Door to ventilation room fixed
- CV
- R-001 & R-002 -> fumehood ventilation was followed up last week, it seems it might not be a ventilation flow rate issue but instead the filters require changing
- R-025 -> the fume cupboard extraction damper opening setpoint must be adjusted directly on the damper using a special device. On the hunt for said special device. At the moment nobody has this adjustment box (if we ever order it from the damper supplier).
- nano alarm -> we know it comes from the pressure in the room not being maintained when the fume cupboard is in use. Tests ongoing.
- target area sas ->delta P not showing correct values, need to follow up and see if there is a connection issue
- MEDICIS GB -> new gauge on order