We describe a multi-anode microchannel plate (MCP) detector with a 16 x 16 discrete pixel array designed for 100 picosecond single photon timing at count rates in the MHz range. The pixel array is instrumented using a 256 channel PETsys TOF ASIC evaluation kit, capable of single photon timing to better than 100 picoseconds and maximum count rate of 640 kHz/channel.
The timing performance of the TOFPET multichannel timing ASIC was measured using electronic stim signals, a Photek PMT210 high speed single anode MCP photomultiplier detector and the multi-anode MCP detector. Both MCP detectors were tested using a pulsed laser operated in single photon counting mode.
The PETsys electronics performance was measured at 43 ps rms using the on-board generated stim pulse synchronous with the TDC clock. MCP detector measurements were made using Illuminated using a pulsed laser with a pulse width of ~40 ps. The time over threshold (ToT) feature of the PETsys ASIC was used to correct the amplitude walk of the due to the MCP pulse height distribution and a single photon timing resolution of better than 100 ps rms was measured.

Primary authors

Presentation Materials

There are no materials yet.
Your browser is out of date!

If you are using Internet Explorer, please use Firefox, Chrome or Edge instead.

Otherwise, please update your browser to the latest version to use Indico without problems.

×