- the budget foresaw some tasks delayed from 2019 - Consolidation of X0 10k - Extra dry gas pipes to X0 10k - Consolidation in gas room 30k - Calibration of 8 flow meters (spares to swap in) - for 2021 - Filters + Misc material + 1/5 of buying 3 maintenance kits 1k - Dryer SX5 annual maintenance (work + sensor recalibration) 2k - Dryer USC55 2k - Dryer 186 (work only) 1k - Operation and Maintenance of gas analyzer 8k (for a student: 16k over 2 years) - Dry Air Sniffer rack maintenance 4k (no Vaisala calibration included) - Misc 2k - Note: from Andromachi: - P5 Bottles - Oil bubbler and switchover system verification - Bld 186 Tasks details from 2019: 1) Gas room: - rack S1J01 supplies detector dry air to cooling bundle flushing (10 lines) and to C6F14 and CO2 cooling cabinets (4), S1J02, S1K02, S1K03 - racks S1J02, S1K02, S1K03 supply from membrane plant to bulkhead flushing (26) --> one rack could be freed up by placing 13 lines in the other two (rack S1K06 supplies from membrane plant to Tracker cold volumes) 2) in X0: - detector dry air from gas room to cooling cabinets runs through X0; --> plan to increase flushing capacity by adding unused pipes in X0 -- although, flow can be increased elsewhere - flushing pixel/strips are supplied by 4 pipes from gas room to X0 racks; --> pixel could be separated from strips by adding one unused pipe in X0 -- although, pixel can get dry after closing sufficiently fast 3) displace PV1 valve normally cutting off S1K06 from membrane plant to also cut off S1K07 -- suggested by Erik 4) GC: main issues are flagged as not being able to ship data to DB and have online measurement, and limited manpower -- no progress since last time 5) Sniffer rack integration into DT supervision system