Evaluating Radiation Safety in SPECT/CT Lymphoscintigraphy: CT Dose Monitoring and Documentation Gaps

20 Jun 2025, 09:30
20m
Gyumri

Gyumri

Speaker

Hassan Salah (Inaya Medical College)

Description

Abstract
This study assesses radiation doses from the CT component of SPECT/CT scans used in lymphoscintigraphy procedures at a single institution. Data from 130 patients were analyzed, focusing on CT dose metrics, including CTDIvol, SSDE, and DLP, alongside technical parameters such as tube current, collimation, and pitch. Results showed an average CTDIvol of 2.07 ± 1.25 mGy (range: 0.08–7.11 mGy), DLP of 81.01 ± 62.15 mGy·cm (range: 3.38–264.58 mGy·cm), and SSDE of 4.97 ± 1.35 mGy (range: 3.14–8.25 mGy). Fixed tube voltage (130 kV) and variable tube current (mean: 65.41 mA, max: 195 mA) were observed, with collimation and pitch tailored to protocols. The CT doses align with diagnostic reference levels, suggesting adherence to safety standards. These findings emphasize protocol optimization and standardized reporting to ensure patient safety while maintaining diagnostic efficacy in hybrid SPECT/CT imaging.
Keywords: SSDE, CTDIvol, DLP, SPECT/CT