Draft:
Present:
Mark, Gunn, James, Chiara, Laurene, Student Stage
Discussion:
- Laurene presented first design of the LHC BGI instrument with 80mm beam pipe and modulated cathode shape
- High voltage feethrough hole too large, can be made smaller by adding ceramic guide
- Big question is chip thermal contact - glue ? Is there acceptance test done by William ?
- Discussion on number of flanges, at the moment there are 4 vacuum seals - anode block, cathode block, HV and chip, can we not reduce the one for chip ?
- Another concern is that the volume around timepix needs discussion as there is also wire bonding step
There are minutes attached to this event.
Show them.