AMBER gas systems
AMBER Gas Systems
Key Outcomes
The team reviewed the current state of gas systems (TPC and detector distribution) and discussed the idea of a multi-year upgrade plan targeting operational readiness for TPC and two other detectors by 2029. A budget request must be submitted by 23 October. The gas systems reconstruction was identified as a high-priority item due to age, lack of spares, and single-point dependency on one expert.
Decisions Made
· Timeline target: TPC gas system must be operational by 2029 (Run 4); parasitic detector tests planned from 2028.
· Other detectors in scope for 2029: MWPCs and GEM detectors only; RICH and other systems excluded for now.
· PLC replacement: Fully replacing the old PLC system/gas systems is mandatory — no spares exist, firmware access is lost, and the sole expert (Stefan) is no longer available.
· Gas circulation strategy: Open systems should move to recirculation to reduce greenhouse gas emissions; aligns with CERN guidelines.
· KT involvement to be explored given the "large" number of systems needed.
TPC-Specific Issues Identified
· Backflow valve→ Andrea: Currently installed normally-closed valve is a safety risk; must be replaced with a normally-open valve to allow safe exhaust.
· Fast evacuation line→ Andrea: No current method to rapidly empty the TPC without risking detector damage; proposed addition of a pressure-reducer + pneumatic valve line to empty within ~1 hour.
· Temperature/cooling→ Andrea?: Gas temperature instability affects density and burst disk safety; proposed solution is a water cooling manifold surrounding the TPC vessel using the existing liquid hydrogen target pump/circuit.
· Compressor oil pressure→ Andrea: Manometers are damaged and unreadable remotely; two pressure transmitters for remote monitoring needed, with auto-stop capability.
· Gas sampling→ Roberto: No certified sampling system exists to collect gas for purity/composition analysis post-run; a certified transport volume must be sourced, with analysis either at CERN (mass spectrometer) or external lab in Lyon (~few hundred euros).
Detector Gas Distribution System Issues
· System designed and built in 2000 by Stefan Berry; documentation incomplete; no spares or firmware access.
· ~6–7 recirculation systems needed in total (MWPCs, GEMs, MicroMegas, W45, new detectors); estimated ~25–30K per unit plus mixer (~15K), totaling potentially 300K+ over the full program.
· Analog-only flow meters in use; shift crew checks manually 3×/day — digital transmitters and remote monitoring needed.
· Filter regeneration frequency unclear; automatic regeneration procedure needed.
Pending Confirmation
· Exact detector volumes, channel counts, and flow rates per system — needed to size recirculation units.
· Whether CERN chilled water integration (via EN-CV) will resolve compressor heat exchanger issues sufficiently.
· KT (Knowledge Transfer) interest in modular recirculation system production.
· Final decision on which open systems must convert to recirculation vs. remain open.
· RICH/C₄F₁₀ system future — currently flushed with argon/nitrogen; decision deferred to Phase 2 planning.
Action Items
· Roberto/Moritz: Deliver overview table of all detector systems (volumes, channel counts, flow rates, contaminants) by end of week.
· Roberto: Provide rough cost estimate for full project (TPC modifications + all recirculation systems + PLC replacement).
· Roberto: Explore KT involvement for knowledge transfer of modular recirculation systems.
· TPC work (modifications): Can be completed in 2027 — sufficient margin before 2029 run start.
Deadline
· Budget request submission: 23 October