Presents: Roland, Pavan, Sebastien, Lukas, Timo and Stefano
- Pavan comments about test report (EDMS 3546993)
- 3 Boris tubes last versions (without glue) are compared :
- without conductive paint
- with conductive paint
- with conductive paint and with triple point filler polyurethane
- each modification improve the electrical behiavour
- Roland: PD defects are unavoidables, if the insulator support the DC HV definded it can be consedered as OK
- in our case, the last variant support 90 kV DC for 10 hours, it can be considered as validated (our nominal is 60 kV)
- Rilis insulator:
- some improvement will be done for the next Rilis insulator:
- the insulator is directly glued on the internal Al tube during wiring procedure by the supplier (no air gap)
- the conductive paint will be applied only on ground side and in the extremities
- we fill it with polyurethane like on the picture (only ground sides)
- the end of the HV sides shall ber rounded like on the picture
- Roland and Pavan aggree to test the DC HV 90kV in their lab, but to be checked if there is enough space inside the cage, if not outside with a "balisage"
- Timo will update the drawings
-