Calorimeter and counting room installed in Pavia's lab
Work with Milano/Insubria group last week to repower the system
FERS:
16 FERS currently mounted and in operation. Modules 409 and 410 (4 FERS) not equipped, because of: (i) lack of LV supply channels; (ii) boards used in Milano for specific tests
Mapping unchanged wrt test beam
12 FERS powered with CAEN Main Frame + 4 (+1: data concentrator) with power adapters
Instabilities observed when powered ON (possibly due to improper firmware initialization) ; it seems solved using slower voltage ramp up
SiPM HV turned ON for all connected channels with no issues
LV boards:
One CAEN A2552 (8 channels) bought new
One CAEN A2519 (8 channels) on loan from Bologna -- cables adapted with D-sub connectors
--> only 4 channels left to PMTs --> ~12 PMTs amplifier boards can be operated
Trigger:
Trigger board in operation. Arduino fw improvements can be tested
2 scintillators used to trigger on cosmics: O(1 Hz), in principle* with low noise (* = no in-depth analysis to see if muons were there)
50 ns total latency between scintillators and FERS --> ok, boards are working in triggered mode
DAQ:
EuDAQ used during TB restarted and working as it was in June
Few k events taken
Fixing priorities, cleaning up, then proceeding with development
Firmware:
Data Concentrator fw upgraded, from 2026.4.1.1 to 2026.11.3.1
FERS fw upgraded, from v7.8 to 9.1
Janus 5 installed on DAQ machine
Link synchronization tested only a few times: ok, no issues (but no evidence of improvements so far, given the very few attempts) --> will be tested as next step
Updated FERSlib to be imported in EuDAQ... debug may be needed…
Status:
System is OFF. We are rearranging the setup in the lab for a more stable configuration
Electromagnetic module:
Equipped with one spare FERS. Data taken with both Janus and EuDAQ
No PMTs. The instrumented 64 channels can be used for calibration studies
14:30
→
14:50
50 Hz noise in triggers20m
Speaker:
Ruggero Turra(Università degli Studi e INFN Milano (IT))
14:50
→
15:10
Fast Simulation of Shower Development in IDEA Detector Dual Readout Calorimeters20m
Speakers: