Speaker
Mr
Roel Van Holen
(Ghent University ELIS-MEDISIP)
Description
INTRODUCTION
Traditional gamma camera imaging with a parallel hole collimated detector (PH) is
limited by the sensitivity versus spatial resolution tradeoff which is intrinsic to
the design. This results in a reduced contrast for small lesions. The higher
geometrical sensitivity of a rotating slat (RS) collimated strip detector could
improve the image quality considerably.
By means of phantom measurements, this study investigates if a RS gamma camera, with
a spatial resolution of 5 mm at 10 cm collimator distance, is able to gain an
improved contrast/noise ratio compared to a traditional camera, equipped with a LEHR
collimator.
METHODS
Projections measured with a RS collimated gamma camera are planar integrals of the
activity distribution and contain less information compared to ray projections,
acquired with a traditional parallel hole collimated gamma camera. Therefore the RS
collimated detecor has to spin around its own axis in order to collect complete data.
Since planar projections are collected at different spin angles, image reconstruction
comparable to a regular SPECT reconstruction is needed to obtain planar images. In
this study we use a Monte Carlo based model of the acquisition physics. Incorporation
of this model in the forward and backward projection of the MLEM algorithm yielded a
Monte Carlo based reconstruction technique (MLEM-MC).
A 17 cm diameter disc of uniform activity was printed on a sheet of paper together
with 12 hot lesions located inside the uniform disc. The lesions had diameters
ranging from 4 to 20mm and a contrast of 4:1. The sheet was placed at 10 cm from the
collimator and 10 different (400 seconds) acquisitions were performed on both the PH
and the RS camera. The data from the RS device were reconstructed in 250 iterations
using MLEM-MC. The images coming from the PH camera were interpolated to have the
same pixel size as the reconstructed RS images (1,8x1,8mm).
The contrast recovery coefficient (CRC), defined as the ratio of the lesion activity
and the background activity, was calculated for each lesion over the 10 different
realisations of the measurement. Afterwards, the mean CRC (mCRC) over all lesions was
calculated. For each pixel, the pixel noise was calculated as the standard deviation
divided by the mean value over the 10 realisations. Averaging over all pixels yielded
a global noise level (NL) of a measurement.
RESULTS
With the NL matched, the RS image reconstructed with MLEM-MC (60 iterations) reached
a 21,0% higher CRC for the smallest hot spot, a 7,6% higher CRC for the 20mm hot
lesion and a mCRC that was 20,0% higher compared to the PH values.
For an equal mCRC, the NL of the RS image was 40,3% lower compared to the PH NL.
For obtaining the same NL on a parallel hole system, the measurement takes 2,81
times longer.
CONCLUSIONS
When the same imaging time is used for all acquisitions, a lower NL is obtained for
RS images at the same mCRC. On the other hand, a higher mCRC can be obtained when
looking at images of the same NL.
For comparable image quality on both modalities, the imaging time could be
significantly reduced using the RS system.
Author
Mr
Roel Van Holen
(Ghent University ELIS-MEDISIP)
Co-authors
Dr
Ignace Lemahieu
(Ghent University ELIS-MEDISIP)
Dr
Stefaan Vandenberghe
(Ghent University ELIS-MEDISIP)
Dr
Steven Staelens
(Ghent University ELIS-MEDISIP)
Dr
Yves D'Asseler
(Ghent University ELIS-MEDISIP)