Speaker
Dr
Antonello Spinelli
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Description
Introduction
The main objective of this work was to characterize the physical performance of a
small animal PET scanner (GE eXplore Vista). The detector modules of the eXplore
Vista consists of an 13 x 13 array of 1.45 mm square “phoswich” elements providing
depth-of-interaction information. Each element is made of two scintillator crystal
layers of Lutetium Yttrium Orthosilicate (LYSO) and Gadolinium Orthosilicate (GSO)
optically glued together in order to form a single 15 mm long segment. The 36
detectors blocks are arranged in two rings giving a total number of phoswich
elements equal to 6084. The transverse field of view (FOV) is equal to 6 cm while
the axial FOV is equal to 4.6 cm.
Typically PET images of small animals are performed using 18-F and 11-C labelled
tracers. Because the mean positron range for the two isotopes is respectively equal
to 0.6 mm and 1.1 mm, it is interesting to investigate the effect of such
difference on the scanner spatial resolution.
Materials and methods
In order to evaluate the performance of the eXplore Vista several parameters such
as spatial resolution, sensitivity, scatter fraction (SF) and noise equivalent
counts (NEC) were measured.
Spatial resolution was measured by using a point source, a line source and a custom
made high resolution phantom. The point source (about 0.2 mm diameter) was obtained
by soaking a small piece of zeolite with a 18-F or 11-C solution. The line source
was obtained by filling a small needle (internal diameter of 0.3 mm) with 18-F or
11-C isotopes. In order to evaluate the dependence of spatial resolution with
respect to transaxial axis position, the line source filled with 18-F was moved
along the y axis and four measurements spaced by 5 mm were acquired. Images of a
custom made high resolution phantom (HRP) with rectangular arrays of holes of 2,
1.5, 1.0 and 0.5 mm with the centres spaced by 4, 3, 2, 1 mm were also acquired.
The HRP was filled with a solution of about 15 MBq of 18-F or 11-C and images were
acquired for 1 h. All the images were reconstructed using Fourier rebinning and
filtered back projection (ramp filter).
In order to measure the absolute eXplore Vista sensitivity a small 18-F point
source of about 1.2 MBq was placed at the centre of the field of view (FOV).
The SF measurements were carried out using a small capillary (0.8 mm internal
diameter) in air and in water (3 cm diameter cylindrical phantom). The SF
measurement was performed accordingly with a method similar as the one described in
the NEMA NU2–2001 report.
In order to estimate the NEC values a rat sized phantom and a mouse sized phantom
were filled with an high amount of 18-F (about 200 MBq) and let to decay for six
half life. Measurements were acquired every 15 minutes.
Results
All the results described in this section are referred to the 250-700 keV energy
window. The transaxial full width half maximum (FWHM) with the point source at the
centre of FOV are respectively equal to 1.48 mm and 1.74 mm for 18-F and 11-C. The
18-F axial resolution is equal to 1.77 mm giving a volumetric resolution of 3.82
mm^3, while for 11-C the axial resolution is 1.91mm and the resulting volumetric
resolution is equal to 5.78 mm^3.
The FWHM measured with the 18-F line source at four transaxial positions were
equal to 1.47 mm, 1.71 mm and 2.1 mm at the centre, 1 cm and 2 cm off-axis,
respectively.
Analysis obtained by drawing several profiles along the HRP hot spots confirmed a
worsening in spatial resolution when using 11-C instead of 18-F (for high
resolution small animal scanners).
The absolute eXplore Vista sensitivity was equal to 2.2% (400-700 keV energy
window), 3.9% (250-700 keV) and 5.9% (100-700 keV) without dead time correction
applied.
For a line source, the measured SF value in air was equal to 19.3%. SF values
obtained in water were equal to 26.2% and 34% for a mouse and a rat-size phantom.
The NEC peak rate were respectively equal to 147 kcps at 710 kBq/ml and 83 kcps at
169 kBq/ml for a mouse and rat-size phantom.
Conclusions
Results show that the eXplore Vista provides high spatial resolution and at the
same time good sensitivity. The value of the spatial resolution for 11-C labelled
tracers was worse respect to 18-F, such loss of resolution for 11-C was also
evident by looking at in vivo mouse images. The eXplore Vista has been in use at
our institution for the last twelve months showing good stability and high
throughput.
Author
Dr
Antonello Spinelli
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Co-authors
Dr
Carlo Bergamini
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Dr
Cinzia Pettinato
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Dr
Cristina Nanni
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Dr
Daniela D'Ambrosio
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Dr
Giuseppe Baldazzi
(Università di Bologna, Bologna, Italy)
Dr
Mario Marengo
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Dr
Silvia Trespidi
(Policlinico S. Orsola-Malpighi, Bologna, Italy)
Dr
Valentina Ambrosini
(Policlinico S. Orsola-Malpighi, Bologna, Italy)