AWAKE spectrometer meeting
Present: Fearghus, Ishkhan, Matthew, Michele, Stefano
Stefano - light calibration
Discussed issues raised in last meeting:
- Flat lamp gives flexibility
- Have some target in front of lamp with e.g. series of points or narrow strip.
- Can also act like LED by having a mask with just one hole.
- Ensure that the distance between screen and lamp is as small as possible.
-> Specifications give relative intensity vertically.
-> Can calibrate each hole (for given mask) using photometer in the lab to get vertical intensity profile.
- Motor controlled remotely. Lamp on/off and intensity can be controlled. These values are recorded via FESA.
-> What about blinkers around the screen ? See if we need to add these around the screen and the light calibration system if needed.
- Order of mirror supports. Stefano signed this to be charged to BI budget; now needs Ans or Edda's approval. If the company receives the order by next week, we will receive the parts by mid September.
Michele - camera software tests
- See slides
Fearghus
- Needs to get DRZ screen cut.
-> To give to Ishkhan on Thursday or next week. Would like for some part of the August run.
- Ordered Basler camera for August run.
-> Check with Ishkhan and Stefano about optics and lens needed.
-> [Stefano] cannot find C <-> F mount converter. Will order one.
- Mirror coating: Ishkhan to check on status.
- Calibration of scintillator response is an issue as we will almost certainly not be able to use the unaccelerated (~15 MeV) electron beam. Fearghus, Matthew and Michele have been discussing this.
- Idea would be to have e.g. X-ray source in AWAKE and measure response to this. Then need to take scintillator and camera to electron test-beam and measure response to electrons and also the X-ray source in this set up.
- Options to investigate for electron beam:
-> Replacement for PHIN injector.
-> CLEAR has electrons ~ 200 MeV.
-> Can we use the calibration lamp instead of a radioactive source to measure the response in the AWAKE area. Do we need an X-ray source or is the lamp sufficient ?
Next meeting, Thursday 3 August at 10:30 / 11:30 (UK / CERN).