Present: Fearghus, Ishkhan, Matthew, Michele, Stefano

 

Stefano - light calibration

Discussed issues raised in last meeting:

- Flat lamp gives flexibility

   - Have some target in front of lamp with e.g. series of points or narrow strip.

   - Can also act like LED by having a mask with just one hole.

- Ensure that the distance between screen and lamp is as small as possible.

-> Specifications give relative intensity vertically.

-> Can calibrate each hole (for given mask) using photometer in the lab to get vertical intensity profile.

- Motor controlled remotely.  Lamp on/off and intensity can be controlled.  These values are recorded via FESA.

-> What about blinkers around the screen ?  See if we need to add these around the screen and the light calibration system if needed.

 

- Order of mirror supports.  Stefano signed this to be charged to BI budget; now needs Ans or Edda's approval.  If the company receives the order by next week, we will receive the parts by mid September.

 

Michele - camera software tests

- See slides

 

Fearghus

- Needs to get DRZ screen cut. 

  -> To give to Ishkhan on Thursday or next week.  Would like for some part of the August run.

- Ordered Basler camera for August run. 

  -> Check with Ishkhan and Stefano about optics and lens needed.

-> [Stefano] cannot find C <-> F mount converter.  Will order one. 

- Mirror coating: Ishkhan to check on status.

- Calibration of scintillator response is an issue as we will almost certainly not be able to use the unaccelerated (~15 MeV) electron beam.  Fearghus, Matthew and Michele have been discussing this.

  - Idea would be to have e.g. X-ray source in AWAKE and measure response to this.  Then need to take scintillator and camera to electron test-beam and measure response to electrons and also the X-ray source in this set up.

  - Options to investigate for electron beam:  

    -> Replacement for PHIN injector.

    -> CLEAR has electrons ~ 200 MeV.

  -> Can we use the calibration lamp instead of a radioactive source to measure the response in the AWAKE area.  Do we need an X-ray source or is the lamp sufficient ?

 

Next meeting, Thursday 3 August at 10:30 / 11:30 (UK / CERN).