ERAM Meeting

→ Europe/Zurich
Alain Delbart (Université Paris-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura-Iuliana Munteanu (CERN)
    • 11:00 → 11:45
      ERAM status
      Conveners: Alain Delbart (Université Paris-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura-Iuliana Munteanu (CERN)
      • 11:00
        ERAM production status at Rui's lab 5m
        Speakers: Alain Delbart (Université Paris-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura Munteanu (CERN), Laura-Iuliana Munteanu (CERN)
      • 11:05
        Mesh pulsing 10m
        Speakers: Alain Delbart (Université Paris-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura Munteanu (CERN), Laura-Iuliana Munteanu (CERN)
        • using FEC#37 + FEC9 & 10 to the mesh pulsing ERAM-35

        mesh pulsing cards FEC 14 & 72 were repaired for damages connectors but had 3+1 dead channels

      • 11:15
        ERAM gluing 5m
        Speakers: Alain Delbart (Université Paris-Saclay (FR)), Charlotte Riccio (CEA-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura-Iuliana Munteanu (CERN)
      • 11:20
        ERAM testing 10m
        Speakers: Alain Delbart (Université Paris-Saclay (FR)), Charlotte Riccio (CEA-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura Munteanu (CERN), Laura-Iuliana Munteanu (CERN)
        • MM1-DLC1 is on the x-ray test bench. Gas connection is done on the 6-7 bar bank
        • The electronics can not be operated as soon as it is plugged on this MM1-DLC1 (power-on and very soon after, the LEDC is switched off)

        debugging is needed : a short-circuit is somewhere (FEC/Stiffener interface ?

        ERAM-02 could be re scanned once it is removed from FC#1

         

      • 11:30
        ERAM AOB 15m
        Speakers: Alain Delbart (Université Paris-Saclay (FR)), Charlotte Riccio (CEA-Saclay (FR)), David Henaff (CEA), Guillaume Eurin (CEA/IRFU/DPhP), Laura-Iuliana Munteanu (CERN)

        shedule of activities (v2.2d, see https://docs.google.com/spreadsheets/d/1ZeinovZhMSxzfOa3VrBpnBEeS8OrXYfAYFpNBGsFN4c/edit#gid=1422499225

        - week 22 (may 29-)

        - ERAM- gluings : first one ERAM-35 --> june 30 ?
        - ERAM-35 x-ray scan
        - FC#1 move in clean room, ERAM removal and storage
        - preparation of the 350V/T2K gas QC setup for ERAMs in clean room
         
        - week 23 (june 5-)
        - FC#2 ERAM testing at 350V in T2K gas (4 ERAMs ?)
        - ERAM-36 gluing ? scan ?
         
        - week 24 (june 12-)
        - FC#2 ERAM testing at 350V in T2K gas (4 ERAMs ?)
        - ERAM-37 gluing ? scan ?
         
        - week 25 (june 12-)
        - Move Bottom TPC in clean room
        - 16 ERAM mounting
         
        - week 26 : possible shift of previous activities