Following our discussion and several internal checks, please find below the agreed plan for the stiffness measurements:
At Room Temperature :
- Tensile tests will be performed on the three supplied cables (2 kA, 18 kA, 5x18 kA and the 4x2 kA cables), each 495 mm in length.
- Measurements will include:
• DIC measurements (distance between the two cable extremities and the clamping systems)
• Crosshead position
- Tests will be carried out up to 7000 N with unloading/loading cycles at 10%, 25%, 50%, 75%, and 100% of the maximum force.
At 77 K :
- Tensile tests will first be repeated at room temperature on one cable (4x2 kA cables) with a reduced length (< 200 mm), using DIC and crosshead position, without the cryostat.
- After this reduced-length room temperature test, the cryostat will be installed, and the test will be repeated without DIC but with crosshead position measurements.
Schedule :
- Room temperature tensile tests are planned for the beginning of next week (Tuesday/Wednesday). Test settings are currently being prepared.
- Depending on your availability to reduce the cable/bundle length, the 77 K test can likely be carried out next Friday or the following Monday.
- Later, all the cable configurations will be tested at 77K.
Additional points :
- To complete the measurement campaign, it would be better to test at least three samples per configuration. Could you please prepare it following the first test scheduled next week?