Speaker
Matthew Gignac
(University of British Columbia (CA))
Description
The instantaneous luminosity of the Large Hadron Collider at CERN will be increased up to a factor of five with respect to the design value by undergoing an extensive upgrade
program over the coming decade. The largest upgrade project for the ATLAS Muon System is the replacement of the present first station in the forward regions with the so-called
New Small Wheels (NSWs), to be installed during the LHC long shutdown in 2018/19. Small-Strip Thin Gap Chambers (sTGC) detectors are one chosen technology to provide fast trigger
and high precision muon tracking under the high luminosity LHC conditions. We study ageing effects of sTGC detectors with a gas mixture of 55% of CO2 and 45% of n-pentane. A sTGC
detector was irradiated with beta-rays from a Sr-90 source. Three different gas flow rates were tested. We observed no deterioration on pulse height of the sTGC up to an accumulated
charge of 2.5 C/cm. By July of this year, we plan to have collected 6 C/cm.
Author
Matthew Gignac
(University of British Columbia (CA))