Description

We have developed a 64 channel ASIC for reading out thousands of SiPM channels for PET applications. A readout electronics based on this ASIC will be described, and we will present the performance of the readout in a test PET scanner setup with 2'048 channels. We will also present a comparison of the performance of our ASIC with SiPMs form different manufacturers First results with a new version of he ASIC will also be presented.

Summary

SiPMs allow a dramatically improvement of PET performance because SiPMs are intrinsically faster than PMTs and because SiPMs can easily be subdivided in a large number of small independent photodetector pixels. The good timing performance of SiPMs will result in better effective sensitivity. The small and independent photodetector pixels allow using one-to-one coupling between a SiPM pixls and a LYSO crystals. This will result in significant improvement of the spatial resolution compared to PMT base systems, where some form of crystal encoding must be used to identify the LYSO crystal where the interaction occurred. Furthermore SiPMs are fundamentally low cost devices. and eventually their price will be lower than the price of PMTs.
A whole body PET scanner will typically have 30'000 LYSO crystals measuring 4x4x20mm. To take advantage of SiPMs in PET applications, it is mandatory to have highly integrated electronics readout. We have developed the 64 channel TOFPET1 ASIC for this purpose. It has 64 independent readout cannels without multiplexing. The output is only digital, 80 bits per event. The rest of the electronics only has to transfer the data to the computer. The coincidence sorting is done in firmware. The readout electronics will scale to many tens of thousands of channels.
The electronics will be described, and we will present the performance of the readout in a test PET scanner setup with 2'048 channels. We will also present a comparison of the performance of our ASIC with SiPMs form different manufacturers
First results with a new version of he ASIC will also be presented.

Primary author

S. Tavernier, Vrije Universiteit Brussel (BE) (Vrije Universiteit Brussel (BE))

Co-authors

Mr Agostino Di Francesco (LIP, Lisbon) Mr Bugalho Ricardo (PETsys electronics, Portugal) Mr Carlos Leong (Petsys electronics, Portugal) Mr Carlos Zorraquino Gaston (Universidad Politécnica de Madrid, Spain) Dr Ferramacho Luis (PETsys electronics, Portugal) Mr Jose Carlos Rasteiro Da Silva (LIP, Lisbon, Portugal) Prof. João Varela (LIP, Lisbon, Portugal) Dr Manuel Rolo (INFN, Torino, Italy) Mrs Niknejad Tahereh (LIP, Lisbon, Portugal) Mr Rui Silva (LIP, Lisbon, Portugal) Dr Silveira Miguel (PETsys electronics, Portugal)