Development of a 1.5 Tesla whole-body MRI Magnet with a very low helium inventory

28 Aug 2017, 13:15
1h 45m
Posters Area

Posters Area

Poster Presentation of 1h45m H1 - Cryostats and Cryogenics Mon-Af-Po1.11

Speaker

Dr Adam Johnstone (Siemens Healthcare)

Description

We present a recent successful program to develop a 1.5 Tesla MRI magnet with a maximum helium inventory of less than 50 litres. The Magnet Technology Demonstrator was fully evaluated in a Siemens MRI system and found to deliver comparable imaging performance to a conventional bath-cooled magnet. As well as the development of the novel magnet technology, other key enabling technologies were developed. These included an automatic ramp-down system for protection against power outages and methods for cooling the magnet both within a manufacturing facility and on site.
The program culminated in the successful demonstration of key technologies required to realise the next generation of whole-body MRI magnets with much reduced dependency on scarce and increasingly expensive helium.

Submitters Country United Kingdom

Primary authors

Dr Adam Johnstone (Siemens Healthcare) Mr Simon Calvert (Siemens Healthcare)

Co-authors

Michael Simpkins (Siemens Healthineers) Simon Chorley (Siemens Healthineers) John Teah (Siemens Healthineers) Jonathan Noys (Siemens Healthineers) Peter Dietz (Siemens Healthineers) Marcel Kruip (Siemens Healthineers) Adrian Bampton (Siemens Healthineers)