Performance evaluation of the developed in-situ ortho-to-parahydrogen fraction measurement system for the ESS cryogenic moderator system

Not scheduled
2h
Posters area

Posters area

Poster Presentation (120m) ICEC 06: Cryogenic applications: hydrogen and LNG systems Wed-Po-2.1

Speaker

HIDEKI TATSUMOTO (European Spallation Source ERIC (ESS))

Description

At the ESS target, high energy spallation neutrons are generated by impinging a 2 MW proton beam on the tungsten target. The proton beam is pulsed with a repetition of 14 Hz and a pulse length of 2.86 ms. The moderator system consists of a water pre-moderator and two liquid hydrogen cold moderators, which are optimized to achieve a high cold neutron brightness. The neutronic performance of the cold moderators degrades rapidly when parahydrogen fraction falls below 99.5%. The neutron collisions in the cold moderators increase the orthohydrogen fraction unless compensated by the ortho- to parahydrogen conversion driven by the catalyst system. Therefore, the Cryogenic Moderator System (CMS) is equipped with an ortho-parahydrogen (OP) catalyst to sustain the desired high parahydrogen fraction. The ortho-to-parahydrogen ratio of liquid hydrogen at the inlet and outlet of the moderators will be measured by the in-situ measurement system (OPMS) using a Raman spectroscopy. A Raman optics system with an accuracy of 0.1% parahydrogen fraction is essential to detect any undesirable shift towards a higher orthohydrogen fraction induced by para-to-ortho back conversion driven by the neutron collisions in the cold moderators. The developed Raman optical probe system has successfully achieved our primary goal of 0.5% accuracy in measurement, as it can detect the fourth parahydrogen peak (J= 4) of normal hydrogen gas at 1240 cm-1.
In this study, Raman spectra of normal hydrogen at ambient temperature were measured while varying hydrogen pressures from 0.10 MPa (0.0827 kg/m3) to 1.1 MPa (0.896 kg/m3). The intensities at J= 0 to 4 were proportional to the hydrogen density. For the pressures higher than 0.7 MPa, the ortho-to-parahydrogen ratios can be measured within an accuracy of ±0.1%. We measured Raman spectra of hydrogen gas passing through a catalyst cooled by liquid nitrogen. The measured ortho-to-parahydrogen ratio was 0.9879±0.0044 and matched closely with the theoretical values of 0.9967. Furthermore, the ortho-to-parahydrogen ratio of hydrogen obtained from the J-PARC CMS under the nominal condition at 21 K was also measured at a pressure of 0.7 MPa and a temperature of 293 K. The parahydrogen fraction was measured to be 99.74 within an accuracy of ±0.11%. It can be expected that the developed Raman optical probe system would achieve the required accuracy of 0.1% for measuring parahydrogen in liquid hydrogen at the ESS CMS, given the significantly higher density of liquid hydrogen.

Submitters Country Sweden

Author

HIDEKI TATSUMOTO (European Spallation Source ERIC (ESS))

Co-authors

Gen Ariyoshi (Japan Atomic Energy Agency) Makoto Teshigawara (Japan Atomic Energy Agency) Takumi Hasegawa (Hiroshima University) Yuuki Shiro (Yamaguchi University) Yuka Horikawa (Yamaguchi University) Hossein Sina (European Spllation Source ERIC (ESS))

Presentation materials