Tue-Mo-Or9 - MRI Magnets I
- Hitoshi Kitaguchi (National Institute for Materials Science)
- Michael Parizh (GE Global Research)
Neurospin is a neuroscience research center located in France at CEA Saclay. The facility is already hosting several MRI magnets and Iseult, an innovative Whole Body 11.7 T MRI system, will be available very soon. The core part of Iseult is an actively shielded NbTi magnet cooled with a superfluid helium bath at 1.8K, that will provide a homogeneous field of 11.7 T within a 90 cm warm bore....
Research and development for the practical application of a medical-use magnetic resonance imaging system (MRI) superconducting magnet that requires without liquid helium started as the New Energy and Industrial Technology Development Organization's (NEDO) supported project in fiscal 2016. Development of a liquid helium-free medical MRI superconducting magnet is desired. Another purpose is to...
Coils for rare-earth barium copper oxide (REBCO) high temperature superconductor (HTS) magnets are often wound with insulation between turns (i.e. effectively infinite resistance). In this case, magnetic field generated by the coil is linearly related to the transport current in the coil, allowing voltage-limited magnet ramping. However, quench protection has been shown to be hard to...
In this study we performed magnetic, mechanical and thermal modeling of a 3T actively shielded whole body MRI (Magnetic Resonance Imaging) magnet. The final design had an overall magnet length and conductor length which can lead to conduction cooled designs which are comparable to NbTi helium bath cooled 3 T designs. The design had a magnetic field homogeneity better than 10 ppm...
In this paper, we present construction and persistent-mode operation results of a half-scale 1.5-T/54-mm room-temperature bore MgB2 MRI-quality magnet for the development of a tabletop “finger” MRI system. A half-scale magnet, composed of 5 coil sections and a persistent-current switch, was wound with a single ~590-m long unreacted/monofilament MgB2 wire having a...
This paper presents the design of a high field and portable MRI magnet system for the rapid and early diagnosis of brain trauma. Early diagnosis and therapy stratification can reduce the risk for critically brain ill patients with the use of near patient imaging, and can aid with precision medicine. High temperature superconductors have the ability to carry large currents at higher...