Alpha-Particle Imaging SiPM Array Sensor with High-Resolution Technique

Not scheduled
20m
Conference room (Wosk Centre)

Conference room

Wosk Centre

Poster Applications in biology, medicine, medical equipments

Speaker

JONGGUK KIM

Description

For a more precise evaluation of the alpha-particle emitting radiopharmaceuticals, images with a high spatial resolution are required. In this study, we propose an alpha-particle imaging SiPM Array Sensor in conjunction with a high-resolution technique. A high-resolution image is recovered using the four images obtained by moving the device along the x- and y-directions. This method uses several low-resolution images acquired by slightly shifting the alpha imaging detector. A high-resolution image is then reconstructed using the MLEM algorithm in conjunction with the measured low-resolution images and the system matrix containing the mechanical shifts, sub-sampling, and measured point spread function of the imaging system. The results show that, using the device with a pre-defined pixel resolution, the proposed technique produces the images of alpha-particles with an improved resolution of approximately 14%. We conclude that the proposed technique has potential in the development of radiopharmaceuticals for targeted alpha therapy.

Submission declaration Original and unplublished

Authors

JONGGUK KIM Dr Kyo Chul Lee (Korea Institute of Radiological and Medical Sciences) Dr kwang Il Kim (Korea Institute of Radiological and Medical Sciences)

Co-author

Dr Hyun Su Lee (Korea Institute of Radiological and Medical Sciences)

Presentation materials

There are no materials yet.