24–26 May 2017
Rayong Marriott Resort & Spa
Asia/Bangkok timezone

Design and construction of portable survey meter

25 May 2017, 17:00
1h
Rayong Marriott Resort & Spa

Rayong Marriott Resort & Spa

http://www.marriott.com/hotels/travel/bkkrr-rayong-marriott-resort-and-spa/ Rayong, Thailand
Poster Instrumentation, Metrology and Standards Poster Presentation II

Speaker

Mr Warut Singseeta (Ubon Ratchathani University)

Description

This work was aimed to design and construction of portable survey meter for radiation dose measuring. The designed system consists of 6 parts including GM-tube radiation detector, low voltage power supply, high voltage power supply, pulse conditioning, processing and display on android phone. The test results indicated that the ripple voltage of low and high voltage power supply were less than 1%. The maximum integral counts were found to be 3×105 cps. It was found that the developed system had small size and light weight for portable measuring system.

Primary author

Mr Warut Singseeta (Ubon Ratchathani University)

Co-authors

Mr Decho Thong-aram (Chulalongkorn University) Dr Somkid Pencharee (Ubon Ratchathani University)

Presentation materials

There are no materials yet.

Peer reviewing

Paper