15–19 Nov 2021
Fukuoka Convention Center
Asia/Tokyo timezone

A Low-Temperature-Dependent Calibration of Hall probes for CPMU

TUE-PO1-115-01
16 Nov 2021, 13:15
2h
Fukuoka Convention Center

Fukuoka Convention Center

Speaker

Dr Chih Wei Chen (National Synchrotron Radiation Research Center)

Description

Abstract
A cryogenic permanent-magnet undulator (CPMU) with a period of 18mm and a magnetic length of 2m is being constructed for the Taiwan Photon Source (TPS). The CPMU with a gap of 5.5 mm can generate an effective magnetic field of 1.2 T at 150 K. When the field measurement, the temperature of hall probe will be drop at cryogenic temperature. Therefore, a field strength calibration and thermostatic temperature control system is necessary. The range of field strength calibration is from 5.5 to ±1.5T at a homemade dipole electromagnet. A 2-axis compact SENIS hall probe is mounted on homemade copper plate in vacuum chamber. The temperature control system consists of a cryocooler、PT100 sensor and a heater to control the Hall probe temperature. Finally, a higher-order polynomial surface fitting to analysis measurement data from calibration system. The field strength maximum error is < 0.2 G at fitting surface. The detailed temperature dependent calibration system is presented in this poster.

Primary author

Dr Chih Wei Chen (National Synchrotron Radiation Research Center)

Co-authors

Yang-Yang Hsu (National Synchrotron Radiation Research Center) Chin-Kang Yang (National Synchrotron Radiation Research Center) Hsiung Chen (National Synchrotron Radiation Research Center) Jui-Che Huang (National Synchrotron Radiation Research Center)

Presentation materials